News Center
We are dedicated to providing users with cost-effective instruments and meticulous services! Welcome to visit, guide, and negotiate business with us.
Current Position:
Home > News > Industrial News

What to do when a thermal gas mass flow meter fails?

2022-07-22
Thermal gas flowmeter is divided into two types of pipeline and insertion type of connection, thermal gas flowmeter has been widely used in the market, our businessmen to buy a lot of thermal gas flowmeter, when we use the thermal gas flowmeter when there is a problem how to solve the problem, the machine malfunction and how to deal with it? Today we talk about how to do when the hot gas flowmeter failure, what are the methods of treatment.

Problems and solutions:
Case 1、No display
1. No power supply 2. Instrument internal switching power supply is damaged 3. DC24V power supply is connected to the opposite 4. display inserted biased 5. display damage
Processing methods
Turn on the power     Turn on the 220VAC power supply, the power indicator does not light up, indicating that the switching power supply is damaged      Detect the polarity of the power supply    Re-insert the screen     ⑤Check the power supply indicator, the indicator light is on, indicating that the screen is damaged, please contact the supplier    


Case 2、Low flow rate
1. Probe direction is reversed 2. Sensor is dirty 3. Sensor is damaged 4. Flow parameters are set incorrectly
Processing methods
Install the probe direction correctly  Clean the sensor  Return to the supplier    Check the parameter setting


Case 3、Abnormal and fluctuating flow rates
1. Flow rate parameter is set incorrectly 2. Fluid nature is pulsating rotation 3. Dirty sensor 4. Damaged sensor
Processing methods
Check the flow rate parameter setting    Adjust the filter coefficient     Clean the sensor      Return to the supplier


Case 4、4-30mA output anomaly
1. 20mA range setting error 2. converter failure 3. wiring not looped
Processing methods
Set the 20mA range correctly  Back to Suppliers    Check wiring


Case 5. Frequency output abnormality
1. Frequency parameters set incorrectly 2. converter is 3. connection line damage Processing methods
Correctly set the frequency parameters   Return to the supplier     Check the connection line


Case 6, The alarm is abnormal
1. Instrument parameters are set incorrectly 2. Instrumentation is not configured alarm output function 3. relay damage
Processing methods
①Correctly set the alarm parameters    Contact the supplier    ③Return to the supplier


Case 7. RS-485 output abnormality
1. The baud rate and station number are set incorrectly 2. Polarity is reversed 3. Connection line is damaged
Processing methods
①Correct input    ②Change polarity     ③Check the connection line
Share To:


Get a Quotation
Please fill out the following form to provide more details about your requirements, contacts, and desired products. We will promptly contact you for further confirmation.