What are the advantages and precautions of the thermal gas mass flow meter?
What are the advantages and precautions of the thermal gas mass flow meter?
a. Pay attention to the working pressure difference, for the instrument to pay special attention to the pressure of the working medium, should pay attention to keep the working pressure difference at both ends of the instrument in and out of the gas within the palliation range. Especially when working under high pressure, the pressure difference is too large, and the flow rate will not be able to be closed or reduced. When using a large flow meter, it is necessary to pay attention to properly thicken the pipeline and reduce the internal resistance of the air source. If the working pressure difference is less than the required value, the flow may not reach the full scale value. When the medium pressure is too high, do not perform online installation and maintenance operations. When the absolute pressure of the measuring medium is higher than 2 standard atmospheres, or higher than the pressure that can cause danger, it should be stopped or depressurized, and the operation should be carried out when the safe pressure is reached.
b. Additional requirements for special media measurements. Some gas medium characteristics are more special, the user needs to specify a special type of product according to the actual situation of the site, before installation, the user should carefully check whether the product type meets the requirements of the site;
c. When the medium is dangerous gas, do not install the box online for maintenance. When the measuring medium belongs to the type of gas that can cause harm to the human body. Online installation and maintenance are prohibited. Relevant security processing should be carried out. Perform the operation only when the site conditions are safe for installation. The online operation without conditions should be discontinued to avoid danger. Such gases as: gas, chlorine, etc.;
d. Disable the flow medium, the use of gas must be purified, avoid dust, liquid and oil. If necessary, install a filter dryer in the gas path. If the instrument outlet is connected with the liquid original bottle, a check valve should be installed at the flow outlet to prevent the liquid backflow from damaging the instrument;
e. Corrosive gases are used. The instrument channel material is 316L stainless steel. General corrosive gases can be measured under the conditions of no water vapor, low leakage, frequent cleaning and proper use in the user system. When used in ammonia, organic solvent vapor or other highly corrosive gases, it should be specially indicated when ordering, so as to replace the sealing material.