The following problems are encountered in the use of thermal gas mass flowmeter
The thermal gas mass flowmeter is particularly widely used and can measure all gases except acetylene, so it is increasingly trusted by users. Many users will encounter some small problems in use, such as:
1. The flow rate is low
a. The probe direction is reversed — the probe direction is installed correctly
b. The sensor is dirty – Clean the sensor
c. Sensor damage – Return to supplier
d. Traffic parameters are incorrectly set — Check parameter Settings
2. The output of the thermal gas mass flowmeter 4-20mA is abnormal
a. Range setting error – correct setting
b. Converter failure — Return to supplier
c. No loop connection — Check the cable connection
3. The RS-485 output is abnormal
A. There is a problem with the baud rate and station number Settings – enter correctly
b. Polarity reversal – change polarity
c. The connection line is damaged — check the connection line