Matters needing attention during the installation of thermal gas mass flowmeter
The thermal gas mass flowmeter is an instrument that uses the principle of heat conduction to measure fluid flow. The instrument uses the constant temperature difference method to measure the gas flow accurately. It has the advantages of small size, high digitization, easy installation and accurate measurement. Today we are mainly to talk about the thermal gas mass flowmeter in the installation process need to pay attention to what are?
1, a type of installation and connection requirements
1) If the converter is installed outdoors, it is necessary to install an instrument box to avoid rain and sun.
2) Prohibit installation in strong vibration occasions, prohibit installation in a large number of corrosive gas environment.
3) Do not share an AC power supply with the equipment that pollutes the power supply such as the frequency converter and welding machine, and if necessary, install a purified power supply for the converter.
4) The one-type insert should be inserted into the axis of the pipeline to be measured, so the length of the measuring rod depends on the diameter of the pipe to be measured, and should be explained when ordering. If it cannot be inserted into the pipe axis, the calibration coefficient will be provided by the manufacturer to complete the accurate measurement.
5) One-type pipe segment is connected with flange, in line with the national standard GB/T9119-2000.
2. Installation of hot gas mass flowmeter
1) One type insert installation by the manufacturer to provide pipe connectors and valves. For non-weldable pipes, pipe fixtures shall be provided by the manufacturer. For example, pipes can be welded. First, the connector is welded to the pipe, then the valve is installed, drilling with special tools, and then the instrument is installed. When maintaining the instrument, remove the instrument and close the valve so as not to affect normal production.
2) Pipe segment installation should select the corresponding standard flange and its connection.
3) When installing, it should be noted that the “medium flow mark” marked on the instrument should be the same as the actual flow direction of the gas.
3. Commissioning and operation of hot gas mass flowmeter
The potential on the instrument enters the measurement state, at this time, the data must be input according to the actual working condition.
4, hot gas mass flow meter maintenance
1) When opening the case, be sure to power off first.
2) When removing the sensor, pay attention to whether the pipeline pressure, temperature or gas is toxic.
3) The sensor is not sensitive to a small amount of dirt, but it should also be cleaned regularly when used in a dirty environment. Otherwise, the measurement accuracy will be affected.以上翻译结果来自有道神经网络翻译(YNMT)· 通用场景